2.2.2.1 Laboratory Dry Matter by Oven Drying for 2 hr at 135°C
Reference:
Drying of feeds at 135oC for 2 hr. (930.15) Official Methods of
Analysis. 1990. Association of Official Analytical Chemists.
15th Edition.
Scope:
This procedure is applicable for the determination of dry matter
on ground air-dry or partially-dried (³85% dry matter) forages
with low volatile acid or alcohol content. Samples dried by this
procedure are not appropriate for further chemical analyses. This
procedure is recommended for developing forage dry matter
calibration for NIR.
Basic Principle:
Moisture is evaporated from the sample by oven drying. Dry matter
is determined gravimetrically as the residue remaining after
drying.
Equipment:
Forced-air drying oven at 135o ±2oC. Oven should be equipped
with a wire rod shelf to allow the circulation of air. It should
be vented and operated with vents open.
Analytical electronic balance, accurate to 0.1 mg Aluminum dish
(pan), ³50 mm diameter, £40 mm deep, covered Desiccator
Reagents:
None.
Safety Precautions:
- Use standard precautions when working around electrical
equipment or glassware.
- Make sure that all electrical equipment is properly
grounded and installed and maintained by qualified
electricians.
Procedure:
- Dry aluminum dish with cover at 135o C ±2o C for at
least 2 hr.
- Cover dish and move to desiccator. Immediately cover
desiccator and allow dishes to cool to room temperature.
Do not allow dishes to remain in desiccator more than 2
to 3 hr.
- Weigh dish with cover (W4) to nearest 0.1 mg, removing
one at a time from desiccator and keeping desiccator
closed between container removals.
- Add approximately 2 g ground sample to each dish. Record
weight of dish with cover and sample (W5) to nearest 0.1
mg.
- Shake dish gently to uniformly distribute the sample and
expose the maximum area for drying.
- Dry samples in oven preheated to 135oC for 2 hr.
- Move samples to desiccator, placing cover on each dish as
it is transferred. Seal desiccator and allow to cool to
room temperature. Do not allow samples to remain in
desiccator for more than 2 to 3 hr.
- Weigh dish with cover and dried sample (W6), recording
weight to nearest 0.1 mg.
For NIR calibration replace step 4 above with:
- 4.1) Load NIR sample cup placing one scoop of
forage (previously dried to 90-95% dry matter or greater,
ground with cyclone mill to pass 1 mm screen, thoroughly
mixed) on each third of the glass surface. Fill to leave
adequate space for cell backing.
- 4.2) Scan sample on NIR instrument and store
spectra.
- 4.3) Immediately remove sample from NIR instrument
and weigh 2 g forage from sample cup to an aluminum dish.
Record weight of dish with cover and sample (W5) to
nearest 0.1 mg.
Comments:
- Time and temperature described in procedures must be
adhered to closely.
- Samples should be placed in drying oven so that air can
circulate freely. Containers should not touch each other.
- Do not overload oven.
- Slide the desiccator lid open. Do not place the lid on
the counter top with the grease side down. The grease
will pick up dirt, preventing formation of a seal.
- Seals should be kept clean and well greased and the lid
should always slide easily on or off. If the lid
"grabs," it is time to remove the old grease
and apply fresh lubricant.
- If a lid can be directly lifted off the desiccator,
either the desiccator was not properly sealed or, more
likely, it needs fresh lubricant.
- Rubber stoppers in the lid should always be pliable.
- Sample dishes should not be packed excessively tight in a
desiccator. Air movement is necessary to cool sample
dishes. Dishes should not touch each other.
- The desiccator lid should be left open for minimal amount
of time.
- Desiccant should be checked and dried periodically.
Replace desiccant twice annually or more often depending
on use. Use of desiccant with color indicator for
moisture is recommended.
- Open a loaded desiccator very slowly after samples have
cooled. A vacuum forms during cooling, and abrupt opening
results in turbulence which can blow samples out of
uncovered containers.
Calculation: Percent Laboratory Dry Matter (Lab DM)
% Lab DM =(W6 - W4/W5 - W4) X 100
- Where W4 = tare weight of dish (with cover) in grams
- W5 = initial weight of sample and dish (with cover) in
grams
- W6 = dry weight of sample and dish (with cover) in grams
Quality Control:
Include at least one set of duplicates in each run if single
determinations are being made.
An acceptable average standard deviation among replicated
analyses for moisture or dry matter is about ±0.10, which
results in a warning limit (2s) of about ±0.20 and a control
limit (3s) of about ±0.30. Plot the results of the duplicate
analyses on an R-control chart (Appendix D) and examine the chart
for trends. Results outside the 95 percent confidence limits warn
of possible problems with the analytical system. Results outside
the 99 percent confidence limits indicate loss of control, and
results of the run should be discarded. If more than five or six
points in succession fall on one side or the other of the 50
percent line, it is a strong indication that something has
changed and is cause for investigation.
|